Permanganate index is reflected in water of organic and inorganic oxidizable substances commonly used indicators. Is defined as: under certain conditions, with certain organic permanganate oxidation of water and inorganic reducing substances, calculated from the amount of potassium permanganate consumed a considerable amount of oxygen.
Features:
The instrument uses GB11892-89 "Quality -Permanganate index measuring the" standard;
Power protection design, with power, then power automatic data recovery;
Intelligent self-diagnostics, equipment management and maintenance very convenient;
Patented design, product failure rate, low maintenance, low reagent consumption;
Suitable for surface water sources and measurements can be applied to sewage Water treatment process testing;
Network-capable, through the network, remote control data can be realized;
Imported PLC control devices to reduce environmental interference and equipment failure;
Using sequential injection analysis, flow path is simple, low reagent consumption, high degree of automation;
Multi-channel valve, strong corrosion resistance, long life, safe and reliable.
Applications:
Chemicals: pharmaceutical, chlor-alkali chemicals, dyes pigments, pulp and paper, intermediates, fertilizers, starch, cornstarch, pesticides, metallurgy, adhesives, detergents and other industries;
Water treatment: air scrubbers, boiler feed water, cooling water, water power plant;
Wastewater: electroplating wastewater, Industrial waste water, heavy metal precipitation.
Technical Parameters:
Model: DH314M1
Methods based on: GB11892-89 "Determination of Water -Permanganate index" - Titration Measuring range: 0 ~ 20mg / L
Accuracy: Indication error ≤ ± 10%
Repeatability error: ≤ ± 5%
Measuring cycle: the minimum measurement period of 45min, according to the actual water samples can be arbitrarily modified 5 ~ 120min digestion time.
Sampling period: time interval (20 ~ 9999min adjustable) and the whole point measurement mode correction: timing instrument sensitivity and zero calibration calibration period: 1 to 99 days at any arbitrary time interval adjustable maintenance cycle: Usually once a month, every views about 30min
Analog output: 4 ~ 20mA
Signal interface: RS232 / RS485
Environmental requirements: Recommended temperature + 5 ~ 28 ℃; humidity ≤90% (non-condensing)
The actual water samples Contrast: Error ≤ ± 10%
MTBF: ≥720h / second Dimensions: 1420 × width 500 × depth 400 (mm)
Power: AC220 ± 10% V, 50 ± 10% Hz, 5A
Other: abnormal alarm and power without losing data; touch-screen display and command input; abnormal reset and power off the call, the instrument automatically discharge the residual reactants is automatically return to work status