Keywords: Environmental protection equipment | Air purifier | Testing Equipment | Chemicals | Valve series
Products
Cyanide Online Automatic monitor
Cyanide Online Automatic monitor

Cyanide leaching of gold is an important industrial solvent gold, most gold producers using the hydrogenation method, and Cyanide is highly toxic and easy to be both a degradation of specialty chemicals. Cyanide comes from a high concentration of industrial wastewater containing cyanide, mainly electroplating wastewater, waste gas cleaning and cooling water cyanide gold extraction, coke oven and blast furnace, some chemical and mineral processing sewage effluent, synthetic rubber, fibers and dyes industry. 

Features: 
Intelligent self-diagnostics, equipment management and maintenance very convenient; 
Error reset, power off the call, the instrument automatically discharge residual reactants and return to work status; 
Automatically determine whether the reagents, beep warning; 
Unique, patented design, product failure rate, low maintenance, low reagent consumption; 
Network-capable, through the network, remote control data can be realized; 
Unique titration system design, with Metering pump flow calibration system to maintain a constant titration accuracy; 
Imported PLC control devices to reduce environmental interference and equipment failure; 
Sequential injection analysis, flow path is simple, low reagent consumption, high degree of automation; 
Multi-channel valve, strong corrosion resistance, long life, safe and reliable.

Technical Parameters: 
Model: DH315C1 
Methods based on: GB / T7486-87 "Water - Cyanide Determination" - potassium persulfate digestion - UV spectrophotometric measurement range: 0 ~ 3.0mg / L (expandable according to customer requirements) 
Accuracy: Indication error ≤ ± 10% 
Repeatability: ≤ ± 10% error 
Measuring cycle: the minimum measurement period 45min 
Sampling period: time interval (10 ~ 9999min adjustable) and the whole point of the measurement mode correction modes: manual and automatic modes of communication output: RS232 and analogue (4 ~ 20) mA 
The actual water samples Contrast: Error ≤ ± 10% 
Dimensions: 1420 × 500 × depth width 400 (mm)

Applications Water treatment plant pollution industrial pollution industrial process water and surface water, groundwater, drinking water sources pollution Water treatment plant

 
点击这里给我发消息